Test Code PROS Protein S Activity, Plasma
Methodology
Amidolysis of Chromogenic Substrate
Performing Laboratory
Huntsville Hospital Laboratory
Specimen Requirements
1. Draw blood in a blue-top (citrate) tube(s). Spin down, remove plasma, spin plasma again, and place 1 mL of citrate platelet-poor plasma in plastic vial. (Glass vial is not acceptable.)
Note: Double-centrifuged specimen
is critical for accurate results as platelet contamination may
cause spurious results.
2. Freeze specimen immediately at ≤-40° C, if
possible.
3. Send specimen frozen.
Note: 1. Each coagulation assay
requested should have its own vial.
2. If the patient is being treated with Coumadin®, this should
be noted.
3. Lipemic specimen may be rejected.
Reference Values
Female: 55-123 % activity
Male: 77-143 % activity
Day(s) Test Set Up
Tuesday, Friday
CPT Coding
85306
Secondary ID
HHA1481
LOINC Coding
Test Name | LOINC Code |
---|---|
Protein S Activity, Plasma |
8265-1 |
Specimen Stability Information
Specimen Type |
Temperature |
Time |
Platelet-poor plasma, citrated |
RT, spun |
4 hrs |
Frozen | 2 weeks |