Test Code (A) PTT Activated Partial Thromboplastin Time (APTT, PTT), Plasma
Methodology
Mechanical Clot Detection
Performing Laboratory
Decatur Morgan Hospital-Decatur Campus
Specimen Requirements
Test must be performed within 4 hours of draw.
Heparinized patients contact Decatur Morgan Hospital-Decatur General Campus Hematology Department at 256-973-2295 prior to draw for special instruction.
1. Draw blood in a blue-top (citrate) tube(s).
2. Spin down, remove plasma, spin plasma again, and place 0.5 mL of
citrate platelet-poor plasma in plastic vial. (Glass vial
is not acceptable.)
Note: Double-centrifuged specimen is critical for accurate results as platelet contamination may cause spurious results.
3. Forward promptly.
Note: Each coagulation assay requested should have its own vial.
Reference Values
22.0-36.0 seconds
Critical value (automatic call-back): ≥150.0
seconds
Test Classification and CPT Coding
85730